Privacy statement: Your privacy is very important to Us. Our company promises not to disclose your personal information to any external company with out your explicit permission.
Select Language
English
This DEC790 Digital Conductivity Sensor is different from the analog sensor, it is a digital Conductivity Sensor with built-in RS485. It uses graphite two-electrode measurement principle. So why use graphite material? Because graphite has good stability, its own conductivity is stable, it is not easily affected by the external environment, and it can work stably for a long time. The conductivity of graphite is also relatively strong in corrosion resistance, and it can work in environments such as acids and alkalis. Graphite itself also has high electrical conductivity, which can provide more accurate measurement results. Most importantly, as a commonly used conductivity sensor, graphite material is low in cost, which can reduce the cost. The graphite is encapsulated in the PPS shell to form our DEC790 digital conductivity sensor.

Generally, the conductivity value of water quality will be affected by temperature, pH value, pressure, etc. In particular, the influence of temperature on conductivity is relatively large. Therefore, this conductivity sensor has a built-in temperature resistance, which can realize automatic temperature measurement and compensation, making the reading more accurate. Because it is a digital sensor, in addition to measuring conductivity, it can also calculate TDS and salinity values. These data can be output through the Modbus communication protocol.
| Product Name | Digital conductivity sensor |
| Signal Type | Digital RS485 Modbus |
| Measuring Range | Conductivity: 0~10mS/cm |
| TDS: 0~5g/L | |
| Salinity: 5ppt(‰) | |
| Temp: 0~50℃ | |
| Accuracy | ± 1% FS |
| Resolution | 0.01μS/cm? ?0.1℃ |
| Response Time | T90<15s |
| Temp. Compensation | Automatic, built-in temperature unit |
| Calibration Method | Multi-point calibration; Contrast Offset; Factor offset |
| Working Pressure | 0~6Bar |
| Operating Temperature | 0~50℃ |
| Shaft Material | PPS |
| Dimensions | Diameter 35mm/Total length 230 mm |
| Lead interface | M16 connector |
| Installation? | 1 inch NPT at both ends; Immersed or piped installation |
| Power Supply | 9~24V DC |
| Protection Grade | IP68 |
?
Privacy statement: Your privacy is very important to Us. Our company promises not to disclose your personal information to any external company with out your explicit permission.
Fill in more information so that we can get in touch with you faster
Privacy statement: Your privacy is very important to Us. Our company promises not to disclose your personal information to any external company with out your explicit permission.